Operational Protocol

The operational protocol of the OPM system is still being written. In the meantime, this page describes the three ways to coregister the head of the participant with the OPM helmet.

First method

  • Laser scan the participant’s head and the stylus points before placing the participant in the OPM helmet

  • Place the participant in the helmet, then laser scan the fiducials on the face again

  • Laser scan the 8 points on the OPM helmet

  • The rigid-body transformation of the head laser scan with respect to the 8 points of the OPM helmet can now be computed from the fiducial laser points

  • This transformation matrix coregisters the laser head scan with the OPM helmet

  • This method assumes that the participant does not move their head within the OPM helmet

  • Advantage: HPI coils are not needed

Second method

  • Laser scan the participant’s head and the stylus points

  • Place the participant in the OPM helmet

  • Place the HPI coils on known stylus points (these locations must be standardized)

  • In this case, a script must be run at the beginning and at the end of the experiment to energize the coils with sinusoidal waves of known frequencies (follow up with the FieldTrip tutorial, section 2)

Third method

  • Laser scan the participant and the fiducials

  • Place the participant in the helmet, then laser scan everything (i.e. the face and the helmet together)

  • Mark the fiducials

  • Coregister both sets of fiducials