Operational Protocol
The operational protocol of the OPM system is still being written. In the meantime, this page describes the three ways to coregister the head of the participant with the OPM helmet.
First method
Laser scan the participant’s head and the stylus points before placing the participant in the OPM helmet
Place the participant in the helmet, then laser scan the fiducials on the face again
Laser scan the 8 points on the OPM helmet
The rigid-body transformation of the head laser scan with respect to the 8 points of the OPM helmet can now be computed from the fiducial laser points
This transformation matrix coregisters the laser head scan with the OPM helmet
This method assumes that the participant does not move their head within the OPM helmet
Advantage: HPI coils are not needed
Second method
Laser scan the participant’s head and the stylus points
Place the participant in the OPM helmet
Place the HPI coils on known stylus points (these locations must be standardized)
In this case, a script must be run at the beginning and at the end of the experiment to energize the coils with sinusoidal waves of known frequencies (follow up with the FieldTrip tutorial, section 2)
Third method
Laser scan the participant and the fiducials
Place the participant in the helmet, then laser scan everything (i.e. the face and the helmet together)
Mark the fiducials
Coregister both sets of fiducials