Operational Protocol
Authors: Gayathri Satheesh gs2750@nyu.edu, Maitha Al Shaali ma6895@nyu.edu, Haidee Paterson haidee.paterson@nyu.edu, Hadi Zaatiti hadi.zaatiti@nyu.edu
Based on a previous version of the protocol from Aniol Santos Angles.
MEG Lab manual DownloadOnce you have mastered this protocol, keep the MEG Session Checklist handy.
Lab booking and schedule
Warning
Do not book on Monday mornings: this is when the helium refill is scheduled, and no data can be acquired during that period. Bookings are also discouraged on Fridays around noon, because the increased movement on campus and in the parking area during prayer time further degrades the quality of your data.
Checks Completed by MEG Scientist Prior to Experiment
The MEG lab is handed over to the project owner after the following checks and tests have been performed successfully.
The KIT system is operational when:
Helium levels are sufficient to conduct an experiment
SQUID sensors are operational
Empty-room data has been acquired, and noise levels have been computed and assessed
The Vpixx system is operational when:
Trigger events are tested
The projector is running
Response boxes are tested for correct button responses
The communication system with the participant is operational when:
The microphone outside the MSR, used by the project owner to talk to the participant, works correctly
The microphone inside the MSR, used by the participant to talk to the project owner, works correctly
The earphones (with disposable single-use foam earplugs), through which the participant hears the project owner, work correctly
The camera inside the MSR, used to monitor the participant, is operational
The laser scanner system is operational when:
The laser scanner computer works correctly
The laser pointer/surface scanner is operational
The eye-tracking system is:
operational
fully connected to the stimulus computer and the MEG-KIT system
The photodiode is:
available in the lab, on the projection screen
operational
Personal protective equipment and clinical consumables are available, including:
clean foam earplugs
head caps
scrubs
clinical application tape for attaching the HPI coils to the participant’s face
gloves
linen
face baby wipes
normal tissue paper
kitchen towels
glasses for vision correction, from -6 to +6 in 0.5 increments
a sign to be used when the participant is a woman wearing a veil/head covering
Should you require new items, please contact the team with your request.
Performed by the project owner - Prepare the lab equipment (prior to participant arrival)
The steps below should be performed by the project owner before the participant arrives, and should be completed in the following order. These procedures take an estimated 20 minutes.
Prepare the MSR
Make sure the MSR has no metal objects inside
Switch the heater off
Heater Button
Prepare bedsheets and pillowcases
Clinical tape is usually stored in the plastic drawers inside the MSR, and also in the top right wooden drawer outside the MSR, on the right side of the Stimulus computer
Prepare 12 pieces of tape; they will be used to attach the HPI coils to the participant’s head
Check the marker box
Ensure that the Marker Box found inside the MSR has enough battery
Power up the Marker Box by flipping the Power switch up
If there is enough battery, the red ‘Low batt’ LED goes on for a second and then back off
If there is not enough battery, the red ‘Low batt’ LED either stays on all the time or never comes on at all. In this case, change the batteries of the box; recharged batteries are available under the Stimulus computer
Marker box power indicator
Uncoil the five HPI marker coils that are linked to the Marker Box
Prepare the Trigger Box
The Trigger Box is outside the MSR and pictured below
Trigger Box found above the MEG Main PC.
Ensure that the Source button is set to LPT1/PC, which is on the left side
If the project owner requires empty-room data prior to the experiment
Turn off the MSR lights and set the light brightness to low
Close the MSR door without a participant inside
On the MEG Main PC computer, open the MEG160 software
Then, Menu -> Acquire -> Auto Tuning -> Ok, and wait for the auto-tuning to be done
From Menu -> Acquire -> MEG Measurement, the Monitor and Acquisition window should open
In the ‘Sensor Control’ window, ensure or set these parameters (NB: these values are only to be used for empty-room data, not for a measurement of neural activity):
HPF to 0.1 Hz
LPF to 1 kHz
BEF to THRU
Click on ‘Sensor Check’
Let the Sensor Check run for around 15 seconds
Uncheck ‘Sensor Check’
Make sure that the sensors display an identical sinusoidal wave on channels 0 to 207
Remember that Sensor 91 is broken and will not display a sine wave
In the ‘Data Acquisition’ window, enter the following:
Patient ID:
sub-emptyroomName:
sub-emptyroom_<date in YYYYMMDD>Foldername:
C:\MEG160\Bin\emptyroom
After ensuring the MSR door is closed, press Lock. The sensor measurements will oscillate rapidly; wait until the values are stable, i.e., no upward or downward trend is observed
Continuous Mode -> Start
Set the Sampling Rate to 2000 Hz
Set the Time to 180 seconds, or to a project-dependent value if you require a different duration of empty-room data
Then, Start Acquisition
When the recording is done, press Unlock
Close the MEG Measurement window
Open the MSR door
Prepare the Vpixx systems
Ensure that the three Vpixx boxes are turned on: Soundpixx, Propixx and Responsepixx
Soundpixx box: all the knobs have to be set, by default, to the positions shown in the image below
Soundpixx Box knob settings
Turn on the computer if it is off, and boot under Windows (it defaults to Ubuntu)
Settings of the Vpixx computer: ensure that the screens are in multiple-display mode (not mirror display), using right-click on the desktop > Display settings > Extend these displays > Keep changes
Set up Vpixx either through the LabMaestro platform (preferred) or through the VPutil command-line tool
LabMaestro
Or use VPutil
Open ‘VPutil’
Type
ppx aand press EnterCheck that the screen inside the MSR is on. If the screen is off, run
ppx s, then runreset, wait for a minute, and runppx aEnsure the orientation (vertical flip) of the screen inside the MSR is correct. If it is not:
Open LabMaestro
On the left of the platform, choose PROPixx from the menu
Check the Rear Projection box
Incorrect screen orientation fix
Verify your experiment script
If using PsychToolBox:
Open MATLAB (2022b)
Open your experiment
.mscript and launch itMake sure you reach the Introduction Page mentioned in the Experiment design section. This ensures that the MEG data recording can be started before the experiment begins
You can do a quick test run to make sure that the trigger signals appear correctly in the MEG160 software
Warning
For a real participant, make sure to turn off the Wi-Fi on the Stimulus Computer so that the experiment is not interrupted by an update or another notification from the internet.
Response device
Button box: make sure all the optical cables from the button boxes are plugged in correctly, as shown in the pictures
Left response box
Right response box
Plugging the response box: the top row is the right response box, the bottom row is the left response box.
Microphone inside the MSR
Make sure the sound box, located under the Stimulus computer, is switched on. If it is not, press the round green button
Check that you can hear, through the speakers, someone talking from inside the MSR into the microphone attached to the dewar
Participant microphone
Earplugs
Check the earplugs and make sure the participant will be able to hear you
Earphones with foam earplugs
Prepare the FastScan computer
If the FastScan computer is not turned on:
Make sure that the FastScan device is off (the flat black box next to the monitor, pictured below)
Turn on the computer, then launch the FastScanII program
Turn on the FastScan device
Fast Scan device
Turn off the doorbell at the entrance of the lab by switching off its plug
Deactivate the doorbell by pressing the plug button.
Perform the MEG Experiment (Participant is present)
If the participant is a woman wearing a veil
Place the sign on the door
Block the door with the isolation screen found behind the laser scan room door
Welcome the participant and explain the session
The script below is a suggestion for what to say at each stage.
[WELCOME] “Thank you for joining our study. Is this your first time in the MEG?”
[GENERAL OVERVIEW] “No worries, let me explain what we are going to do today.”
[BEFORE MEG - HEAD SHAPE] “Before you go into the MEG, we need to do some preparation.” Then explain the FastScan head laser scan:
“I will scan the shape of your head with a laser gun [show the FastScan]. This gives us a 3D reconstruction of the shape of your head. To do that, you need to sit there without moving for around 5 to 7 minutes.”
“I also have to mark five points on your forehead and close to your ears with this washable ink [show it]; it will disappear after just one shower [show the phantom head with the points].”
“Why are we doing that? To know where your head is located while you are in the MEG. This is important for the study we are running, because we need to know where the data recorded by the MEG sensors, which measure the tiny changes in the magnetic field generated by brain activity, is coming from. Different people have different head shapes and sizes, and they place their head in slightly different positions relative to the MEG sensors.”
“Why the points? When we are in the MEG room, I will tape small things called ‘head position coils’ on the places where you have these painted points, and this will tell us where your head is relative to the MEG sensors. It looks a bit weird at the beginning, but you get used to it quickly (I did the experiment on myself).”
[BEFORE MEG - CLOTHES] “Another important thing is that you cannot go inside the MEG room with any kind of metallic object, because it will create an artifact on the MEG sensors. To ensure that, I have to ask you to wear these MEG-compatible clothes (like the ones in hospitals). Please, if you feel comfortable with that, take off your bra (most of the time there are small metallic strips or parts).”
[INSIDE MEG] Explain the study-specific instructions here, or give the participant an instruction manual to read. “Now, let me recap what we will do today. You need to fill in the forms, we scan the shape of your head, then you change clothes. You go to the MEG room, we tape the coils on your forehead, and then you will do the tasks.”
[END OF EXPLANATION] “Is everything clear? Do you have any questions? Do you feel comfortable? Are you OK? Please let me know; it is important for us that you understand everything.”
Fill in the forms
Ensure that all consent forms, and any forms needed specifically for your experiment, have been signed electronically, or allow the participant to fill them in and sign them by hand at the time of the session
Check for MEG incompatibilities
Participants MUST change from their street clothes into the provided scrubs. Underwear without any metal and socks can be kept on. Bras should be removed
Make sure they have NO metallic objects in the body/eyes
Confirm that they do not have:
Surgical clips, artificial heart valve, implanted drug pump
Bullet
Cochlear implant or hearing aid
Dental retainers or braces
Make-up, especially red-coloured make-up
Hair pins
Jewelry and piercings
Keys
Phone
If the participant arrives with make-up, ask them to remove it completely
If the participant wears glasses, ask them to remove them and provide them with MR-compatible glasses from the briefcase found in the lab. Determine their vision prescription and give them the closest matching pair of glasses from the briefcase (available lens strengths for vision correction: -6 to +6 in 0.5 increments)
MR safe glasses briefcase.
Ask the participant to put their phone on airplane mode
Put your own phone and all other phones in the MEG lab on airplane mode
Call the security guard on 85849 and ask them to turn off their walkie-talkies for the duration of the experiment
Phone in MEG lab with a label of security guard office.
Perform the FastScan laser head scan
Cap the participant
Put a swimming cap on the head of the participant
Make sure the cap is as smooth as possible on the participant’s head
Participants with long hair can keep most of their hair outside the cap, behind their ears and down their back
The ears must be clear of hair
The cap must cover all the hair that can be seen at the anterior, left and right parts of the head
Smooth the hair under the cap as much as possible; excess long hair can be drawn outside the cap at the cerebellum level
The goal is for the cap to follow the shape of the skull as closely as possible
Mark the fiducials (there are 8 in total)
Use the “T” template, with the line aligned with the participant’s nasion as in the picture below
“T” template on the left and nasion/pre(auricular) positions on the right
Mark the nasion using a pen (fiducial 1)
Adjust the fold in the “T” template to the participant’s nasion
Using a pen marker, mark fiducials 6, 7 and 8 through the three holes in the “T” template
Fiducials numbered by the order they should be laser scanned with.
Mark the left and right tragi and the left and right pre-auricular regions (1 cm anterior to the tragi): points 2, 3, 4 and 5
For sanitary purposes, place a tissue on the neck brace over the area closest to the mouth (see picture)
Put on the neck brace
Neck brace with tissue for sanitary purposes
Perform the laser scan, once FastScan has finished initializing (indicated at the bottom of the software UI)
Open the FastScan II software on the computer
Press ‘New’
Ensure the scanner is in Sweep mode
Ask the participant to close their eyes and avoid any movement until the scan is finished
Point the laser gun at the laser scanner reference receiver (the box on the ring you place around the neck, see below) with a half-click, followed by a full click
Laser scanner reference receiver
Neck brace with laser scanner reference point on the bottom left
Warning
Error message during scanning
If the following error message appears, the laser scan device has lost the reference point.
LaserScan error when device is de-referenced
Do not press Cancel on the message. Point the laser scanner at the reference receiver, first with a half click to point and then with a full click.
Scan the head shape (sweeps) with a full click. Tips:
Include ALL cap surfaces and the surfaces with fiducial points
Avoid overlapping sweeps
Make separate sweeps for the head and the face
Keep a consistent distance between the head and the scanner throughout the scan
Hint
Press half a click while using the laser scanner to shift the view in the FastScan II software to the current view as seen from the device. This feature allows you to quickly identify areas that are not well covered by the current laser scan.
Register the fiducial points
After the sweeps, switch to Laser Points and click on Stylus List for the points options; ensure that Stylus > Properties > Capture Points is selected (NOT capture lines)
Register the 8 fiducial points in the following order (see the fiducials picture above):
Nasion - between the eyebrows at the nasal bridge
Left tragus - cartilage of the left ear (not marked)
Right tragus - cartilage of the right ear (not marked)
Left marker - pre-auricular marking, left ear
Right marker - pre-auricular marking, right ear
Center forehead - center forehead marking
Left forehead - left forehead marking
Right forehead - right forehead marking
Ensure that you have only 8 points selected in the Stylus List
Tell the participant that they can move again
Put the scanner away
Return the scanner and box to the foam holder on the table, and make sure none of the cords are on the floor. THIS IS A VERY EXPENSIVE DEVICE (see picture)
FastScan should ALWAYS be placed like this: laser on foam, cables on table (not floor).
Remove the cap from the participant’s head and toss it into the washing bin
Remove the neck brace
Files must be saved and stored according to the BIDS format, either at this point or after the session is done; refer to the next sections for saving instructions
The participant goes into the MSR
The participant should remove their shoes
The participant sits on the bed facing the researcher
Place the five Head Position Indicator (HPI) coils on the marker points
Each HPI coil is marked with a color that corresponds to its position on the head
Bring the forehead markers over the top of the head so that the wires are not in the participant’s face
The position of each HPI coil on the participant’s head should follow the mapping below
Note that the red and yellow markers should be placed anterior to the tragus, at marked points 4 and 5
Position of Marker Coils also called Head Position Indicator (HPI) coils
Place the earphones
Assist the participant with the wires while they move into the helmet
Place the pink pillow under the participant’s legs for back support
If the experiment requires it, place the Vpixx Response Box or the Dial in their right or left hand, depending on the experimental design. Optionally tape the box to the mattress to prevent it from falling
Tape any loose wires of the markers and the button boxes
Ensure that the participant is comfortable
Close and lock the MSR door
HPI Coils placement on head.
Communicate with the participant
Turn on the microphone
Talk to the participant through the Vpixx microphone
Make sure the participant replies and that the voice quality is good
Tell them that the experiment is about to start and that they should refrain from any movement
Tell them that if they need to speak to you about any urgent issue, they can do so freely at any time
Turn off the microphone
Run the experiment script
Run your script until it reaches its Introduction Page, as explained in the Experiment design section
Then prepare the MEG recording as described in the next step
Prepare the MEG recording
On the ‘MEG MAIN PC’ computer, open MEG Lab (on the desktop), also known as MEG160
The participant is now in the MSR, the door is CLOSED, the lights have been switched off and the brightness dial has been turned to the minimum
If empty-room data has not been acquired prior to the current experiment, Auto Tuning should be performed
Menu -> Acquire -> Auto Tuning -> Ok
Wait for the auto-tuning to be done
This should be followed by a Sensor Check
From Menu -> Acquire -> MEG Measurement, the Monitor and Acquisition window should open
In the ‘Sensor Control’ window, ensure or set these parameters:
HPF to 0.1 Hz
LPF to 500 Hz
BEF to THRU
Click on ‘Sensor Check’
Let the Sensor Check run for around 30 seconds
Uncheck ‘Sensor Check’
Make sure that the sensors display an identical sinusoidal wave
Remember that Sensor 91 is broken and will not display a sine wave
File storage location and naming convention
From the menu “Acquire (Q)”, select “MEG Measurement (Q)”
In the “Monitor and Acquisition” window > ‘Data Acquisition’, enter:
Patient ID:
sub-<subjectID>[e.g.,sub-001]Patient Name:
sub-<subjectID>_ses-<session-number>_task-<projectname>[e.g.,sub-001_ses-01_task-resting-state]Foldername:
D:\MEGDATA\<Lab_name>\STUDY_CODE\sub-<subjectID>
“Lock” [only if the MSR door is CLOSED]
Wait until the MEG sensors are stable, i.e. no upward or downward trend
Perform a marker measurement
Switch off the microphone
In the “Monitor and Acquisition” window: Marker measurement > Start > OK (see picture)
When done, the ‘GOF%’ column should be around 99%
If not, at least one of the head coils is misplaced
If at least 3 head marker coils are in place, PROCEED
Click OK
Continuous mode (left) and Marker measurement (right).
A
.mrkfile namedYYYYMMDD-x.mrkis automatically generated in the specified directory after the marker measurement, wherexis an integer (1, 2, 3, …) indicating the order in which the markers were recordedIf your experiment is lengthy (for example 2 hours long), we recommend that you perform a marker measurement in between, for example after 1 hour has elapsed
In the “Monitor and Acquisition” window:
Set or ensure that:
HPF: 0.1 Hz
LPF: 500 Hz
BEF: THRU
Continuous Mode > Start (see the picture above)
Sampling rate: 1000 (default)
Time: 4000 seconds [about 66 minutes] (this is the maximum possible time in the MEG160 software)
Start Acquisition
At this point, you can safely start your experiment from the Stimulus computer
If your experiment is lengthy, we recommend that you perform a marker measurement between runs
A marker measurement must be performed at the end of an experiment
Stop the continuous recording (when the task finishes, or if the experiment lasts for more than 4000 seconds and needs a new recording)
In the “Monitor and Acquisition” window (see the picture above): Continuous Mode > Abort
Finish the MEG session (when all tasks are done!)
In the “Monitor and Acquisition” window:
‘Unlock’ [VERY IMPORTANT STEP, DO NOT OPEN THE DOOR BEFORE IT]
Close the MEG160 software
Take the participant out of the MSR
[ONLY WHEN THE SENSORS ARE UNLOCKED!] Open the MSR door
Remove the head-position indicator coils and the earphones yourself. The coils in particular are very fragile and expensive; remove them with care
Ask the participant to change back into their clothes and to put the scrubs in the wash bin (in the laser scan room)
After the MEG session
MEG settings
Do not shut down any of the computers. They can all be locked or logged off
Turn the heater back on by pressing the heater button again [THIS IS VERY IMPORTANT] (see the picture above)
Turn off the MSR lights by pressing the switch-off button and turning the light dimmer to the minimum
Double-check that you turned the heater back on
Clean the room
Remove the linen and place it in the bin
Clean the helmet, head-position indicator coils and button box with wipes
Wipe down the FastScan neck brace and any other surfaces the participant came in contact with
Postprocessing
Apply the Noise Reduction filter using the reference magnetometers
The KIT system is equipped with reference magnetometers on channels 208 to 223, which measure the external magnetic field
[Optional] You can noise-reduce your SQUID data (channels 0 to 207) by applying a filter that uses the data from the reference channels (we only use channels 208, 209 and 210, which measure the ambient magnetic field along three orthogonal axes, x, y and z):
Open the produced
.confile in the default app MEG160, then apply a Noise Reduction filter using Edit -> Noise ReductionMake sure the magnetometers on channels 208, 209 and 210 are used
Execute the noise reduction, then File -> Save As and add
_NRat the end of the file nameTransfer both files to NYU Box, as detailed in the data uploading section
FastScan instructions
Open your
.fsnfile if it is not already open; the FastScanII software should launchFile -> Save As, and save the
.fsnfile in the desktop folderLaser Scan Data\[Your project name]Then go to Edit > Generate Surface
Smoothing = 5 mm
Decimation = 3 mm
Surface Simplification = 0.10
File > Export > Save as, basic surface: first save the surface file in the same folder, then press Yes to export the stylus points file to the same folder as well
For the BIDS naming convention, please refer to MEG-Laserscan files
Uploading to NYU Box
You should have your own folder on NYU Box, named after your project
Refer to the Data uploading section to structure your BIDS directory and upload your data
Appendix A: Stylus location and markers
The following table summarises each registered stylus location and whether a KIT marker coil is placed at that position.
Index |
Body Part |
KIT marker coil |
|---|---|---|
1 |
Nasion |
No |
2 |
Left Tragus |
No |
3 |
Right Tragus |
No |
4 |
Left Pre-Auricular |
Yes |
5 |
Right Pre-Auricular |
Yes |
6 |
Center Forehead |
Yes |
7 |
Left Forehead |
Yes |
8 |
Right Forehead |
Yes |
Appendix B: Order of appearance of the KIT marker coils in the .mrk file
The registered .mrk file contains the positions of the HPI coils for KIT.
Using the FieldTrip function ft_read_headshape('PATH TO .mrk'), we report below the order of appearance
of the HPI coil positions in the .mrk file.
This has been tested with many .mrk files in the current plugging setting (last column).
Order of appearance in the .mrk |
Placing position of HPI Coil on head |
Color |
Plugging order in Marker Box |
|---|---|---|---|
1 |
Central Forehead (CF) |
Blue |
2 |
2 |
Left Pre-Auricular (LE) |
Red |
0 |
3 |
Right Pre-Auricular (RE) |
Yellow |
1 |
4 |
Left Forehead (LF) |
White |
3 |
5 |
Right Forehead (RF) |
Black |
4 |