Operational Protocol

Authors: Gayathri Satheesh gs2750@nyu.edu, Maitha Al Shaali ma6895@nyu.edu, Haidee Paterson haidee.paterson@nyu.edu, Hadi Zaatiti hadi.zaatiti@nyu.edu

Based on a previous version of the protocol from Aniol Santos Angles.

MEG Lab manual PDF version of this protocol Download

Once you have mastered this protocol, keep the MEG Session Checklist handy.

Lab booking and schedule

Warning

Do not book on Monday mornings: this is when the helium refill is scheduled, and no data can be acquired during that period. Bookings are also discouraged on Fridays around noon, because the increased movement on campus and in the parking area during prayer time further degrades the quality of your data.

Checks Completed by MEG Scientist Prior to Experiment

The MEG lab is handed over to the project owner after the following checks and tests have been performed successfully.

  1. The KIT system is operational when:

    • Helium levels are sufficient to conduct an experiment

    • SQUID sensors are operational

    • Empty-room data has been acquired, and noise levels have been computed and assessed

  2. The Vpixx system is operational when:

    • Trigger events are tested

    • The projector is running

    • Response boxes are tested for correct button responses

  3. The communication system with the participant is operational when:

    • The microphone outside the MSR, used by the project owner to talk to the participant, works correctly

    • The microphone inside the MSR, used by the participant to talk to the project owner, works correctly

    • The earphones (with disposable single-use foam earplugs), through which the participant hears the project owner, work correctly

    • The camera inside the MSR, used to monitor the participant, is operational

  4. The laser scanner system is operational when:

    • The laser scanner computer works correctly

    • The laser pointer/surface scanner is operational

  5. The eye-tracking system is:

    • operational

    • fully connected to the stimulus computer and the MEG-KIT system

  6. The photodiode is:

    • available in the lab, on the projection screen

    • operational

  7. Personal protective equipment and clinical consumables are available, including:

    • clean foam earplugs

    • head caps

    • scrubs

    • clinical application tape for attaching the HPI coils to the participant’s face

    • gloves

    • linen

    • face baby wipes

    • normal tissue paper

    • kitchen towels

    • glasses for vision correction, from -6 to +6 in 0.5 increments

    • a sign to be used when the participant is a woman wearing a veil/head covering

Should you require new items, please contact the team with your request.

Performed by the project owner - Prepare the lab equipment (prior to participant arrival)

The steps below should be performed by the project owner before the participant arrives, and should be completed in the following order. These procedures take an estimated 20 minutes.

  1. Prepare the MSR

    • Make sure the MSR has no metal objects inside

    • Switch the heater off

      Heater Button Image

      Heater Button

    • Prepare bedsheets and pillowcases

    • Clinical tape is usually stored in the plastic drawers inside the MSR, and also in the top right wooden drawer outside the MSR, on the right side of the Stimulus computer

    • Prepare 12 pieces of tape; they will be used to attach the HPI coils to the participant’s head

  2. Check the marker box

    • Ensure that the Marker Box found inside the MSR has enough battery

      • Power up the Marker Box by flipping the Power switch up

      • If there is enough battery, the red ‘Low batt’ LED goes on for a second and then back off

      • If there is not enough battery, the red ‘Low batt’ LED either stays on all the time or never comes on at all. In this case, change the batteries of the box; recharged batteries are available under the Stimulus computer

      Marker box power indicator

      Marker box power indicator

    • Uncoil the five HPI marker coils that are linked to the Marker Box

  3. Prepare the Trigger Box

    • The Trigger Box is outside the MSR and pictured below

      Trigger Box

      Trigger Box found above the MEG Main PC.

    • Ensure that the Source button is set to LPT1/PC, which is on the left side

  4. If the project owner requires empty-room data prior to the experiment

    • Turn off the MSR lights and set the light brightness to low

    • Close the MSR door without a participant inside

    • On the MEG Main PC computer, open the MEG160 software

    • Then, Menu -> Acquire -> Auto Tuning -> Ok, and wait for the auto-tuning to be done

    • From Menu -> Acquire -> MEG Measurement, the Monitor and Acquisition window should open

      • In the ‘Sensor Control’ window, ensure or set these parameters (NB: these values are only to be used for empty-room data, not for a measurement of neural activity):

        • HPF to 0.1 Hz

        • LPF to 1 kHz

        • BEF to THRU

      • Click on ‘Sensor Check’

      • Let the Sensor Check run for around 15 seconds

      • Uncheck ‘Sensor Check’

      • Make sure that the sensors display an identical sinusoidal wave on channels 0 to 207

      • Remember that Sensor 91 is broken and will not display a sine wave

    • In the ‘Data Acquisition’ window, enter the following:

      • Patient ID: sub-emptyroom

      • Name: sub-emptyroom_<date in YYYYMMDD>

      • Foldername: C:\MEG160\Bin\emptyroom

    • After ensuring the MSR door is closed, press Lock. The sensor measurements will oscillate rapidly; wait until the values are stable, i.e., no upward or downward trend is observed

    • Continuous Mode -> Start

      • Set the Sampling Rate to 2000 Hz

      • Set the Time to 180 seconds, or to a project-dependent value if you require a different duration of empty-room data

      • Then, Start Acquisition

    • When the recording is done, press Unlock

    • Close the MEG Measurement window

    • Open the MSR door

  5. Prepare the Vpixx systems

    • Ensure that the three Vpixx boxes are turned on: Soundpixx, Propixx and Responsepixx

    • Soundpixx box: all the knobs have to be set, by default, to the positions shown in the image below

      Soundpixx box settings

      Soundpixx Box knob settings

    • Turn on the computer if it is off, and boot under Windows (it defaults to Ubuntu)

    • Settings of the Vpixx computer: ensure that the screens are in multiple-display mode (not mirror display), using right-click on the desktop > Display settings > Extend these displays > Keep changes

    • Set up Vpixx either through the LabMaestro platform (preferred) or through the VPutil command-line tool

    • LabMaestro

      • Open LabMaestro, found on the desktop

        LabMaestro_icon

        LabMaestro icon

      • To turn on the PROPixx screen, click on the ‘Wake All PROPixx’ button found in the menu at the top of the LabMaestro platform

        LabMaestro Platform

        LabMaestro Platform

    • Or use VPutil

      • Open ‘VPutil’

      • Type ppx a and press Enter

      • Check that the screen inside the MSR is on. If the screen is off, run ppx s, then run reset, wait for a minute, and run ppx a

      • Ensure the orientation (vertical flip) of the screen inside the MSR is correct. If it is not:

        • Open LabMaestro

        • On the left of the platform, choose PROPixx from the menu

        • Check the Rear Projection box

        Incorrect screen orientation fix

        Incorrect screen orientation fix

  6. Verify your experiment script

    • If using PsychToolBox:

      • Open MATLAB (2022b)

      • Open your experiment .m script and launch it

      • Make sure you reach the Introduction Page mentioned in the Experiment design section. This ensures that the MEG data recording can be started before the experiment begins

    • You can do a quick test run to make sure that the trigger signals appear correctly in the MEG160 software

    Warning

    For a real participant, make sure to turn off the Wi-Fi on the Stimulus Computer so that the experiment is not interrupted by an update or another notification from the internet.

  7. Response device

    • Button box: make sure all the optical cables from the button boxes are plugged in correctly, as shown in the pictures

    Left Button Box

    Left response box

    Right Button Box

    Right response box

    Response box plugs

    Plugging the response box: the top row is the right response box, the bottom row is the left response box.

  8. Microphone inside the MSR

    • Make sure the sound box, located under the Stimulus computer, is switched on. If it is not, press the round green button

    • Check that you can hear, through the speakers, someone talking from inside the MSR into the microphone attached to the dewar

      Participant microphone

      Participant microphone

  9. Earplugs

    • Check the earplugs and make sure the participant will be able to hear you

      Earphones with foam earplugs

      Earphones with foam earplugs

  10. Prepare the FastScan computer

    • If the FastScan computer is not turned on:

      • Make sure that the FastScan device is off (the flat black box next to the monitor, pictured below)

      • Turn on the computer, then launch the FastScanII program

      • Turn on the FastScan device

      Fast Scan device

      Fast Scan device

  11. Turn off the doorbell at the entrance of the lab by switching off its plug

    Doorbell

    Deactivate the doorbell by pressing the plug button.

Perform the MEG Experiment (Participant is present)

  1. If the participant is a woman wearing a veil

    • Place the sign on the door

    • Block the door with the isolation screen found behind the laser scan room door

  2. Welcome the participant and explain the session

    The script below is a suggestion for what to say at each stage.

    • [WELCOME] “Thank you for joining our study. Is this your first time in the MEG?”

    • [GENERAL OVERVIEW] “No worries, let me explain what we are going to do today.”

    • [BEFORE MEG - HEAD SHAPE] “Before you go into the MEG, we need to do some preparation.” Then explain the FastScan head laser scan:

      • “I will scan the shape of your head with a laser gun [show the FastScan]. This gives us a 3D reconstruction of the shape of your head. To do that, you need to sit there without moving for around 5 to 7 minutes.”

      • “I also have to mark five points on your forehead and close to your ears with this washable ink [show it]; it will disappear after just one shower [show the phantom head with the points].”

      • “Why are we doing that? To know where your head is located while you are in the MEG. This is important for the study we are running, because we need to know where the data recorded by the MEG sensors, which measure the tiny changes in the magnetic field generated by brain activity, is coming from. Different people have different head shapes and sizes, and they place their head in slightly different positions relative to the MEG sensors.”

      • “Why the points? When we are in the MEG room, I will tape small things called ‘head position coils’ on the places where you have these painted points, and this will tell us where your head is relative to the MEG sensors. It looks a bit weird at the beginning, but you get used to it quickly (I did the experiment on myself).”

    • [BEFORE MEG - CLOTHES] “Another important thing is that you cannot go inside the MEG room with any kind of metallic object, because it will create an artifact on the MEG sensors. To ensure that, I have to ask you to wear these MEG-compatible clothes (like the ones in hospitals). Please, if you feel comfortable with that, take off your bra (most of the time there are small metallic strips or parts).”

    • [INSIDE MEG] Explain the study-specific instructions here, or give the participant an instruction manual to read. “Now, let me recap what we will do today. You need to fill in the forms, we scan the shape of your head, then you change clothes. You go to the MEG room, we tape the coils on your forehead, and then you will do the tasks.”

    • [END OF EXPLANATION] “Is everything clear? Do you have any questions? Do you feel comfortable? Are you OK? Please let me know; it is important for us that you understand everything.”

  3. Fill in the forms

    • Ensure that all consent forms, and any forms needed specifically for your experiment, have been signed electronically, or allow the participant to fill them in and sign them by hand at the time of the session

  4. Check for MEG incompatibilities

    • Participants MUST change from their street clothes into the provided scrubs. Underwear without any metal and socks can be kept on. Bras should be removed

    • Make sure they have NO metallic objects in the body/eyes

    • Confirm that they do not have:

      • Surgical clips, artificial heart valve, implanted drug pump

      • Bullet

      • Cochlear implant or hearing aid

      • Dental retainers or braces

      • Make-up, especially red-coloured make-up

      • Hair pins

      • Jewelry and piercings

      • Keys

      • Phone

    • If the participant arrives with make-up, ask them to remove it completely

    • If the participant wears glasses, ask them to remove them and provide them with MR-compatible glasses from the briefcase found in the lab. Determine their vision prescription and give them the closest matching pair of glasses from the briefcase (available lens strengths for vision correction: -6 to +6 in 0.5 increments)

      Glasses briefcase

      MR safe glasses briefcase.

    • Ask the participant to put their phone on airplane mode

    • Put your own phone and all other phones in the MEG lab on airplane mode

    • Call the security guard on 85849 and ask them to turn off their walkie-talkies for the duration of the experiment

      Phone

      Phone in MEG lab with a label of security guard office.

  5. Perform the FastScan laser head scan

    • Cap the participant

      • Put a swimming cap on the head of the participant

      • Make sure the cap is as smooth as possible on the participant’s head

      • Participants with long hair can keep most of their hair outside the cap, behind their ears and down their back

      • The ears must be clear of hair

      • The cap must cover all the hair that can be seen at the anterior, left and right parts of the head

      • Smooth the hair under the cap as much as possible; excess long hair can be drawn outside the cap at the cerebellum level

      • The goal is for the cap to follow the shape of the skull as closely as possible

    • Mark the fiducials (there are 8 in total)

      • Use the “T” template, with the line aligned with the participant’s nasion as in the picture below

        Template and Nasion

        “T” template on the left and nasion/pre(auricular) positions on the right

      • Mark the nasion using a pen (fiducial 1)

      • Adjust the fold in the “T” template to the participant’s nasion

      • Using a pen marker, mark fiducials 6, 7 and 8 through the three holes in the “T” template

        Fiducials

        Fiducials numbered by the order they should be laser scanned with.

      • Mark the left and right tragi and the left and right pre-auricular regions (1 cm anterior to the tragi): points 2, 3, 4 and 5

      • For sanitary purposes, place a tissue on the neck brace over the area closest to the mouth (see picture)

      • Put on the neck brace

        Neck brace with tissue for sanitary purposes

        Neck brace with tissue for sanitary purposes

    • Perform the laser scan, once FastScan has finished initializing (indicated at the bottom of the software UI)

      • Open the FastScan II software on the computer

      • Press ‘New’

      • Ensure the scanner is in Sweep mode

      • Ask the participant to close their eyes and avoid any movement until the scan is finished

      • Point the laser gun at the laser scanner reference receiver (the box on the ring you place around the neck, see below) with a half-click, followed by a full click

        Reference point

        Laser scanner reference receiver

        Neck brace with laser scanner reference point

        Neck brace with laser scanner reference point on the bottom left

        Warning

        Error message during scanning

        If the following error message appears, the laser scan device has lost the reference point.

        LaserScan error when device is de-referenced

        LaserScan error when device is de-referenced

        Do not press Cancel on the message. Point the laser scanner at the reference receiver, first with a half click to point and then with a full click.

      • Scan the head shape (sweeps) with a full click. Tips:

        • Include ALL cap surfaces and the surfaces with fiducial points

        • Avoid overlapping sweeps

        • Make separate sweeps for the head and the face

        • Keep a consistent distance between the head and the scanner throughout the scan

        Hint

        Press half a click while using the laser scanner to shift the view in the FastScan II software to the current view as seen from the device. This feature allows you to quickly identify areas that are not well covered by the current laser scan.

    • Register the fiducial points

      • After the sweeps, switch to Laser Points and click on Stylus List for the points options; ensure that Stylus > Properties > Capture Points is selected (NOT capture lines)

      • Register the 8 fiducial points in the following order (see the fiducials picture above):

        1. Nasion - between the eyebrows at the nasal bridge

        2. Left tragus - cartilage of the left ear (not marked)

        3. Right tragus - cartilage of the right ear (not marked)

        4. Left marker - pre-auricular marking, left ear

        5. Right marker - pre-auricular marking, right ear

        6. Center forehead - center forehead marking

        7. Left forehead - left forehead marking

        8. Right forehead - right forehead marking

      • Ensure that you have only 8 points selected in the Stylus List

      • Tell the participant that they can move again

    • Put the scanner away

      • Return the scanner and box to the foam holder on the table, and make sure none of the cords are on the floor. THIS IS A VERY EXPENSIVE DEVICE (see picture)

        Fast Scanner Box

        FastScan should ALWAYS be placed like this: laser on foam, cables on table (not floor).

      • Remove the cap from the participant’s head and toss it into the washing bin

      • Remove the neck brace

      • Files must be saved and stored according to the BIDS format, either at this point or after the session is done; refer to the next sections for saving instructions

  6. The participant goes into the MSR

    • The participant should remove their shoes

    • The participant sits on the bed facing the researcher

    • Place the five Head Position Indicator (HPI) coils on the marker points

      • Each HPI coil is marked with a color that corresponds to its position on the head

      • Bring the forehead markers over the top of the head so that the wires are not in the participant’s face

      • The position of each HPI coil on the participant’s head should follow the mapping below

      • Note that the red and yellow markers should be placed anterior to the tragus, at marked points 4 and 5

      Colour coded marker placement

      Position of Marker Coils also called Head Position Indicator (HPI) coils

    • Place the earphones

    • Assist the participant with the wires while they move into the helmet

    • Place the pink pillow under the participant’s legs for back support

    • If the experiment requires it, place the Vpixx Response Box or the Dial in their right or left hand, depending on the experimental design. Optionally tape the box to the mattress to prevent it from falling

    • Tape any loose wires of the markers and the button boxes

    • Ensure that the participant is comfortable

    • Close and lock the MSR door

      hpi coils

      HPI Coils placement on head.

    • Communicate with the participant

      • Turn on the microphone

      • Talk to the participant through the Vpixx microphone

      • Make sure the participant replies and that the voice quality is good

      • Tell them that the experiment is about to start and that they should refrain from any movement

      • Tell them that if they need to speak to you about any urgent issue, they can do so freely at any time

      • Turn off the microphone

  7. Run the experiment script

    • Run your script until it reaches its Introduction Page, as explained in the Experiment design section

    • Then prepare the MEG recording as described in the next step

  8. Prepare the MEG recording

    • On the ‘MEG MAIN PC’ computer, open MEG Lab (on the desktop), also known as MEG160

    • The participant is now in the MSR, the door is CLOSED, the lights have been switched off and the brightness dial has been turned to the minimum

    • If empty-room data has not been acquired prior to the current experiment, Auto Tuning should be performed

      • Menu -> Acquire -> Auto Tuning -> Ok

      • Wait for the auto-tuning to be done

    • This should be followed by a Sensor Check

    • From Menu -> Acquire -> MEG Measurement, the Monitor and Acquisition window should open

      • In the ‘Sensor Control’ window, ensure or set these parameters:

        • HPF to 0.1 Hz

        • LPF to 500 Hz

        • BEF to THRU

      • Click on ‘Sensor Check’

      • Let the Sensor Check run for around 30 seconds

      • Uncheck ‘Sensor Check’

      • Make sure that the sensors display an identical sinusoidal wave

      • Remember that Sensor 91 is broken and will not display a sine wave

    • File storage location and naming convention

      • From the menu “Acquire (Q)”, select “MEG Measurement (Q)”

      • In the “Monitor and Acquisition” window > ‘Data Acquisition’, enter:

        • Patient ID: sub-<subjectID> [e.g., sub-001]

        • Patient Name: sub-<subjectID>_ses-<session-number>_task-<projectname> [e.g., sub-001_ses-01_task-resting-state]

        • Foldername: D:\MEGDATA\<Lab_name>\STUDY_CODE\sub-<subjectID>

      • “Lock” [only if the MSR door is CLOSED]

      • Wait until the MEG sensors are stable, i.e. no upward or downward trend

    • Perform a marker measurement

      • Switch off the microphone

      • In the “Monitor and Acquisition” window: Marker measurement > Start > OK (see picture)

      • When done, the ‘GOF%’ column should be around 99%

      • If not, at least one of the head coils is misplaced

      • If at least 3 head marker coils are in place, PROCEED

      • Click OK

        Daq measurement

        Continuous mode (left) and Marker measurement (right).

      • A .mrk file named YYYYMMDD-x.mrk is automatically generated in the specified directory after the marker measurement, where x is an integer (1, 2, 3, …) indicating the order in which the markers were recorded

      • If your experiment is lengthy (for example 2 hours long), we recommend that you perform a marker measurement in between, for example after 1 hour has elapsed

    • In the “Monitor and Acquisition” window:

      • Set or ensure that:

        • HPF: 0.1 Hz

        • LPF: 500 Hz

        • BEF: THRU

      • Continuous Mode > Start (see the picture above)

      • Sampling rate: 1000 (default)

      • Time: 4000 seconds [about 66 minutes] (this is the maximum possible time in the MEG160 software)

      • Start Acquisition

      • At this point, you can safely start your experiment from the Stimulus computer

    • If your experiment is lengthy, we recommend that you perform a marker measurement between runs

    • A marker measurement must be performed at the end of an experiment

  9. Stop the continuous recording (when the task finishes, or if the experiment lasts for more than 4000 seconds and needs a new recording)

    • In the “Monitor and Acquisition” window (see the picture above): Continuous Mode > Abort

  10. Finish the MEG session (when all tasks are done!)

    • In the “Monitor and Acquisition” window:

      • ‘Unlock’ [VERY IMPORTANT STEP, DO NOT OPEN THE DOOR BEFORE IT]

      • Close the MEG160 software

  11. Take the participant out of the MSR

    • [ONLY WHEN THE SENSORS ARE UNLOCKED!] Open the MSR door

    • Remove the head-position indicator coils and the earphones yourself. The coils in particular are very fragile and expensive; remove them with care

    • Ask the participant to change back into their clothes and to put the scrubs in the wash bin (in the laser scan room)

After the MEG session

  1. MEG settings

    • Do not shut down any of the computers. They can all be locked or logged off

    • Turn the heater back on by pressing the heater button again [THIS IS VERY IMPORTANT] (see the picture above)

    • Turn off the MSR lights by pressing the switch-off button and turning the light dimmer to the minimum

    • Double-check that you turned the heater back on

  2. Clean the room

    • Remove the linen and place it in the bin

    • Clean the helmet, head-position indicator coils and button box with wipes

    • Wipe down the FastScan neck brace and any other surfaces the participant came in contact with

  3. Postprocessing

    • Apply the Noise Reduction filter using the reference magnetometers

      • The KIT system is equipped with reference magnetometers on channels 208 to 223, which measure the external magnetic field

      • [Optional] You can noise-reduce your SQUID data (channels 0 to 207) by applying a filter that uses the data from the reference channels (we only use channels 208, 209 and 210, which measure the ambient magnetic field along three orthogonal axes, x, y and z):

        • Open the produced .con file in the default app MEG160, then apply a Noise Reduction filter using Edit -> Noise Reduction

        • Make sure the magnetometers on channels 208, 209 and 210 are used

        • Execute the noise reduction, then File -> Save As and add _NR at the end of the file name

        • Transfer both files to NYU Box, as detailed in the data uploading section

  4. FastScan instructions

    • Open your .fsn file if it is not already open; the FastScanII software should launch

    • File -> Save As, and save the .fsn file in the desktop folder Laser Scan Data\[Your project name]

    • Then go to Edit > Generate Surface

      • Smoothing = 5 mm

      • Decimation = 3 mm

      • Surface Simplification = 0.10

    • File > Export > Save as, basic surface: first save the surface file in the same folder, then press Yes to export the stylus points file to the same folder as well

    • For the BIDS naming convention, please refer to MEG-Laserscan files

  5. Uploading to NYU Box

    • You should have your own folder on NYU Box, named after your project

    • Refer to the Data uploading section to structure your BIDS directory and upload your data

Appendix A: Stylus location and markers

Markers MEG-system image Markers MEG-system image

The following table summarises each registered stylus location and whether a KIT marker coil is placed at that position.

Index

Body Part

KIT marker coil

1

Nasion

No

2

Left Tragus

No

3

Right Tragus

No

4

Left Pre-Auricular

Yes

5

Right Pre-Auricular

Yes

6

Center Forehead

Yes

7

Left Forehead

Yes

8

Right Forehead

Yes

Appendix B: Order of appearance of the KIT marker coils in the .mrk file

The registered .mrk file contains the positions of the HPI coils for KIT. Using the FieldTrip function ft_read_headshape('PATH TO .mrk'), we report below the order of appearance of the HPI coil positions in the .mrk file. This has been tested with many .mrk files in the current plugging setting (last column).

Order of appearance in the .mrk

Placing position of HPI Coil on head

Color

Plugging order in Marker Box

1

Central Forehead (CF)

Blue

2

2

Left Pre-Auricular (LE)

Red

0

3

Right Pre-Auricular (RE)

Yellow

1

4

Left Forehead (LF)

White

3

5

Right Forehead (RF)

Black

4