System specification

System Overview

The OPM system was installed on 4 March 2024.

Fieldline OPM system

Fieldline OPM system with 96 sensors, installed at NYUAD.

Powering on or off the OPM

The OPM racks must always be turned on or off starting from the top rack.

Sensor locations on helmet

The sensor locations on the helmet are shown in the image /usr/share/hedscan/doc/Beta2SensorLocations.png on the OPM computer. In addition, every saved .fif file contains the location of the channels in (x, y, z) coordinates.

Trigger channels on OPM

OPM to Vpixx digital pins connection

A custom cable has been built to link the Vpixx system to the OPM system. The pin mapping is given below; the cable datasheet is available in the internal version of this page.

Pin Mapping

Pin

Sender: DB25 Digital Output (female port) VPIXX

Receiver: DB37 Digital Input (male port) OPM

1

Dout0

Data Bit 0 (pin2)

2

Dout2

Data Bit 2 (pin4)

3

Dout4

Data Bit 4 (pin6)

4

Dout6

Data Bit 6 (pin8)

5

Dout8

Data Bit 8 (pin11)

6

Dout10

Data Bit 10 (pin13)

7

Dout12

Data Bit 12 (pin15)

8

Dout14

Data Bit 14 (pin17)

9

Dout16

Data Bit 16 (pin20)

10

Dout18

Data Bit 18 (pin22)

11

Dout20

Data Bit 20 (pin24)

12

Dout22

Data Bit 22 (pin26)

13

Ground

Ground (pin1)

14

Dout1

Data Bit 1 (pin3)

15

Dout3

Data Bit 3 (pin5)

16

Dout5

Data Bit 5 (pin7)

17

Dout7

Data Bit 7 (pin9)

18

Dout9

Data Bit 9 (pin12)

19

Dout11

Data Bit 11 (pin14)

20

Dout13

Data Bit 13 (pin16)

21

Dout15

Data Bit 15 (pin18)

22

Dout17

Data Bit 17 (pin21)

23

Dout19

Data Bit 19 (pin23)

24

Dout21

Data Bit 21 (pin25)

25

Dout23

Data Bit 23 (pin27)

OPM sensors and direction of measurement

In its current state, the OPM system measures the radial magnetic field (Bz) but not yet the tangential directions (Bx or By). The hardware is capable of measuring both, but FieldLine has not yet released the update that enables the tangential directions to be measured.

Dynamic Range

The dynamic range of a sensor is the ratio between the highest amplitude it can measure linearly, before reaching saturation, and its noise floor, below which a signal cannot be distinguished from noise.

\[D = \frac{B_\text{max}}{B_\text{noise}}\]

It is often expressed in decibels (dB) as

\[D_\text{dB} = 20 \, \log_{10} \!\left( \frac{B_\text{max}}{B_\text{noise}} \right).\]

To compute the dynamic range of an MEG sensor:

  • measure the highest amplitude \(B_\text{max}\) that the sensor can measure, for example by applying a slowly varying (DC) source and observing the measurement

  • measure the noise level \(B_\text{noise}\), for example by recording empty-room data and computing the average level of the noise (a measured signal can be distinguished from the sensor noise only when its amplitude is above this level)

OPM HPI Coils

The HPI coils are placed on specific landmarks on the head of the participant.

The 6 HPI coils are connected from the OPM system into the MSR through 6 brass cables with the following characteristics:

  • TNP 3.5mm Mono Extension (25FT) - 12V Trigger

  • IR Infrared Sensor Receiver Extension Extender

  • 3.5mm 1/8” TS Monaural Mini Mono Audio Plug Jack Connector Male to Female Cable Wire Cord

  • Length: 7.6 meters

A script must be run before starting the experiment to activate the HPI coils. The script energises the HPI coils with known sinusoidal waves (see Appendix D of the OPM documentation).

The script can be run at the beginning and at the end of the experiment to register the coil positions, which are then averaged. Coregistration strategies are discussed in the Operational Protocol.